Small drybox:
- MBraun LabStar glovebox filled with argon, 1 work-stand;
- constant dynamic circulation with automatically regenerated gas purification system with removal of moisture and oxygen/oxides;
- circuit feeded from gas cylinder with 99.999% argon (5N) with <1ppm H2O content;
- 2 vacuum antechambers (small one and big one) connected to high-performance oil vacuum pump (possibility of obtaining vacuum up to 0.1 Pa);
- total oxygen-containing impurities (moisture/oxygen/oxides) content inside the glovebox is smaller than 0.5ppm;
- ceramic total oxygen analyzer;
- there are magnetic stirrer, coin cell assembly post and digital analytical balance (d=0.0001g) inside the glovebox;
- possibility of performing electrochemical measurements inside the glovebox (adequate cable leads available);
- fully automated and programmable pumping of the large antechamber;
- HEPA filters on the inlet and outlet of the circulation system (atmosphere inside free from small particles);
|
|
Smaller drybox:
- glovebox filled with argon, 1 work-stand;
- constant dynamic circulation with gas purification system with removal of moisture;
- circuit feeded from gas cylinder with 99.999% argon (5N) with <1ppm H2O content;
- vacuum antechamber connected to oil vacuum pump (possibility of obtaining vacuum up to 0,1 Pa);
- total impurities content inside the glovebox is smaller than 50ppm, work with pure lithium metal possible;
- there are magnetic stirrers inside the glovebox;
- possibility of performing electrochemical measurements inside the glovebox (adequate cable leads available);
|
|
Bigger drybox:
- glovebox filled with argon, 4 work-stands;
- constant dynamic circulation with gas purification system with removal of moisture and oxygen;
- circuit feeded from gas cylinder with 99.999% argon (5N) with <1ppm H2O content;
- 2 vacuum antechambers (small one and big one) connected to oil vacuum pump with molecular sieves (possibility of obtaining vacuum up to 10 Pa);
- moisture content inside the glovebox is smaller than 25ppm (because of work with solvents)
- electrochemical moisture probe;
- there are magnetic stirrers and digital analytical balance (d=0.0001g) inside the glovebox;
- possibility of performing electrochemical measurements inside the glovebox (adequate cable leads available);
|
|